PH/EC/Temp meter MW803 milwaukee

Product no.: MW803


It can be shipped within 2 days

130.00 / unit(s)

Price includes VAT and delivery to Spain



The PH / EC / TDS / Temp MW803 meter gives you accurate results, it has a screen that shows the results and it is very light.


 

MW803 PH/EC/TDS/Temperature Tester with Replaceable Electrode

  • MW803 is a portable device with a two-level LCD display that measures pH/EC/TDS/Temperature in a single tester.
  • The large display shows values ​​in the extended range of 0.00 to 14.00 pH and 0 to 3999 μS/cm, 0 to 2000 ppm (MW803), 0 to 20.00 mS/cm, 0 to 10.00 ppt ( MW804) and simultaneously displays the temperature. 0.0 to 50.0°C or 32.0 to 122.0°F.
  • They have a stability indicator and a "HOLD" function that freezes the screen to make it easier to read and record measured values.
  • The large screen also has graphic symbols to guide you through all operations.
  • The EC/TDS conversion factor is user adjustable, as is the temperature compensation factor (β).
Ideal for quick and accurate measurements in swimming pools, aquariums and gardening applications. They can also be used in industrial and laboratory applications, such as cooling towers, food processing, plating, wastewater and drinking water treatment, etc.

Specifications:

  • Measurement range: pH 0.00 to 14.00, EC 0.0 to 3999 mS/cm, TDS 0 to 2000 ppm, temperature 0.0 to 50.0 °C
  • Sensitivity: pH 0.01, EC 1 mS/cm, TDS 1 ppm, temperature 0.1°C
  • Deviation: pH ± 0.05, EC/TDS ± 2% of full scale, Temp ± 0.5°C
  • pH electrode: Mi60P (replaceable)
  • Applicable in the environment: 0 to 50°C; maximum relative humidity 100%
  • Batteries: 4 x 1.5 V, IEC LR44, A76 (included)
  • Battery life: approx. 100 hours of continuous operation
  • Dimensions: 254 x 67 x 47mm
  • Weight: 220g
  • Calibration: automatic 1 step for EC, 1 to 2 steps for pH
  • Automatic temperature compensation with β = 0.0 to 2.4%/°C

Combo PH/EC/TDS/Temp Meter MW803

  • Operation of the MW803 PH/EC/TDS/Temp Combo Meter:
  • Remove protection;
  • Turn on the MW803 by pressing the ON/OFF Button;
  • Immerse the electrode in the liquid to measure.
  • The MW803 automatically compensates temperature in PH/EC/TDS.
  • To turn off the meter, press the ON/OFF/CAL button.

Measurement with the MW803 PH/EC/TDS/Temp Combo Meter:

  • To choose the PH measurement mode, either EC (µS), or TDS (ppm), press the SET HOLD key to switch from one parameter to another.
  • Once the measurement parameter has been chosen, measure by immersing the electrode in two cm of liquid to be measured.
  • To measure another parameter, continue using the SET HOLD key.

Calibration of the MW803 PH/EC/TDS/Temp Combo Meter:

  • Turn on the meter,
  • Using the SET HOLD key, choose the parameter to calibrate (PH, or µS, or ppm), and
  • Press the ON/OFF/CAL button until the OFF message is replaced by the CAL message.

The meter is in calibration mode for the chosen parameter. PH Calibration:

  • The message “pH7.01 use” appears on the screen, asking to be immersed in pH7.01 calibration liquid.
  • The meter will indicate “OK” and will ask to be immersed in pH4.01 liquid.
  • After being immersed in pH4.01 liquid, after a few seconds, it will indicate “OK2” which means it is calibrated at 2 points, and it will automatically return to PH measurement mode.

EC Calibration (µS):

The message “1413µS” appears on the screen, asking to be immersed in 1413uS calibration liquid.
The meter will indicate “OK” once calibrated, and will automatically return to EC measurement mode.

TDS Calibration (ppm):

  • The message “1382ppm” appears on the screen, asking to be immersed in 1382ppm calibration liquid.
  • The meter will indicate “OK” once calibrated, and will automatically return to TDS measurement mode.

Electrode change:

  • Unscrew the white ring,
  • Remove the Mi60P electrode and change it with a new one,
  • Be careful not to damage the connections and without forgetting to place the black rubber on the new electrode.

IMPORTANT for the MW803 PH/EC/TDS/Temp Combo Meter: Never use distilled or deionized water as storage liquid; Also do not use 90º alcohol

We also recommend

*

Price includes VAT and delivery to Spain


Browse these categories as well: Meters pH, chlorine and others, Complete catalogue